Change Language Home Log In Shopping Cart Sales & Service Site Map Contact Us
Fluid System Technologies Search  
 

Swagelok® CR-288® Concentration Monitor
The Swagelok CR-288 concentration monitor provides real-time, temperature-compensated concentration measurement in a compact, ultrahigh-purity package. The monitor uses an optical system to accurately measure liquid chemical concentration in real time. The digital display unit (DDU) and 288-connect® software package combine convenient scalable data interface and acquisition options via a graphical user interface and liquid crystal display.

Concentration monitor, CR-288

Sensor Specifications
Fluid Pressure Rating
Fluid Temperature Rating
Process Fluid Index of Refraction Range
Body Material
Optical Window

DDU Specifications
Power Requirements
Power Supply Input


0 to 80 psig (0 to 5.5 bar)
60 to 149°F (16 to 65°C)
1.3138 to 1.4108

DuPont® Teflon® (modified PTFE)
Semiconductor-grade sapphire

24 V(dc) ± 10 %; 0.5 A max
100/240 V (ac); 50/60 ± 3 Hz


Semiconductor International Editors' Choice Best Product 2008

Swagelok CR-288 named Semiconductor International Editors' Choice
Best Product 2008

More Information
Download the Concentration Monitor—CR-288 catalog.
Download the Swagelok CR-288 Concentration Monitor User's Manual.
Contact your authorized Swagelok sales and service representative.



Concentration Monitor Features
  • Provides process monitoring for inline liquid chemical applications
  • Provides real-time information for point-of-use chemical mixing/blending, spiking and dilution without process intrusion or interruption
  • PC-based graphical user interface software for data collection, analysis, and field calibration.

Concentration sensor, CR-288Sensor

  • Compact, ultrahigh-purity package
  • Bowl design optimizes fluid flow over sensor
  • Meets SEMI F57-0301
  • Fine thread flare and Nippon Pillar® Super 300 end connections standard; other end connections available



CR-288 Software288-Connect Software

  • Windows® based
  • Interface into real-time process
  • Real-time concentration and temperature data logging
  • Allows rapid field calibration to specific process fluid



Concentration monitor cableCable

  • Connects the sensor to the digital display unit
  • FEP jacketed, EMC-shielded 10 ft (3 m) instrumentation cable
  • IP67-rated snap-on connector to HD 15 serial connector



Concentration monitor digital display unitDigital Display Unit

  • LCD display
  • 4 to 20 mA output for temperature and concentration
  • Two 5 V digital output channels
  • RS-232 communication with computer software package
  • Connects up to four sensors per DDU
  • RS-232 serial DB25/DB9 to USB 1.1 adapter included
  • White powder-coated aluminum housing with mounting slots



Performance Data

  • Refractive Index Accuracy (Refractive Index Units, RIU): ± 1 × 10–4
  • Refractive Index Repeatability: 1 × 10–5
  • Concentration Accuracy: ± 0.1 % 1
  • Concentration Repeatability: 0.01 %1
  • Temperature Accuracy: ± 0.1°C
  • Temperature Resolution and Repeatability: 0.01°C
  • Response Time: 1.2 s (undamped)
  • Damping is user configurable.

1 Based on measuring ethylene glycol and water at 20 psig (1.3 bar), 76.1 to 78.8°F (24.5 to 26.0°C), and a calibration range of 0 to 100 % concentration.


Cleaning and Packaging
Every Swagelok CR-288 sensor is cleaned and packaged in accordance with SEMI Standard F57-301 for ultrahigh-purity fluid system components.


Testing and Calibration
Every Swagelok CR-288 sensor is factory leak tested to meet SEMI Standard E49.3, Section 9.2 for allowable leak rate. Each sensor is factory calibrated to deionized water at 20°C. The sensor calibration table can be updated with the 288-connect software to optimize performance with other liquids.


Traceability

Sensor bonnets are laser marked for traceability.



Find a Sales & Service Center